CYNOSURE LUTRONIC
Lutronic DermaV KTP 532 nm and Nd:YAG 1064 nm laser for vascular and pigmented lesions
Two wavelengths in a single platform to treat facial veins, rosacea, angiomas and pigmented spots on the face and body.
WHY CHOOSE IT
LUTRONIC DERMAV KEY BENEFITS
1
Two lasers, one platform
532 and 1064 nm in the same device: choose the right wavelength for each lesion without switching devices.
2
Fast on large areas
IntelliTrak adapts energy delivery to the handpiece speed and reduces the number of passes needed.
3
Patient comfort
ICD cryogen cooling and real-time temperature monitoring; minimal downtime.
DermaV is manufactured by Cynosure Lutronic. Bioskin Italia distributes it and supports every installation with clinical training, technical service and marketing support.
THE PRODUCT
What is Lutronic DermaV?
Lutronic DermaV is a dermatology laser platform that combines two sources in a single device: a 532 nm KTP green laser and a long-pulsed 1064 nm Nd:YAG laser. It is designed for dermatologists and aesthetic physicians who treat vascular and pigmented lesions and want to switch wavelengths within the same session, without changing devices.
THE TECHNOLOGY
How it works: two wavelengths, two depths
DermaV works by selective photothermolysis: the laser light is absorbed by a specific target, haemoglobin in the vessels or melanin in pigmented spots, which heats up and coagulates. The surrounding tissue absorbs little of that wavelength and stays protected.
532 nm KTP: superficial vessels and pigment
Green 532 nm light is strongly absorbed by haemoglobin and melanin and works at the surface. It is the choice for facial veins, rosacea, diffuse redness and lentigines.
1064 nm Nd:YAG: deeper vessels and darker skin types
1064 nm penetrates deeper and is poorly absorbed by epidermal melanin. It is used for larger, deeper vessels such as venous lakes and leg veins, and allows darker skin types to be treated more safely.
IntelliTrak and temperature control
A sensor in the handpiece detects the scanning speed and adjusts energy delivery, keeping coverage uniform even on large areas such as legs, back and décolletage. Skin temperature is monitored in real time, and ICD (Intelligent Cooling Device) cryogen cooling protects the epidermis during each pulse. DermaV is the first device to combine cryogen cooling with the 532 and 1064 nm wavelengths.
Tailored pulses
With Variable Sequential Pulsing, DermaV works in single-pulse, sub-milli and sub-micro modes, with pulse durations from 0.1 to 60 ms: according to Lutronic, it covers treatments that used to require pulsed dye and long-pulsed lasers. The IntelliTone, Clear & Smooth and ART (Accelerated Rejuvenation with Tracking) protocols address tone, texture and rejuvenation.
INDICATIONS
What DermaV treats
Vascular lesions
- Telangiectasia and facial veins
- Couperose and erythematous rosacea
- Angiomas and haemangiomas
- Port-wine stains
- Venous lakes
- Reticular veins and leg veins
Benign pigmented lesions
- Solar and age-related lentigines
- Café-au-lait macules
- Seborrhoeic keratoses
For more resistant pigmented lesions, Bioskin also offers Lutronic PICO Plus and Lutronic Hollywood Spectra.
Other indications
- Stable long-term hair reduction with 1064 nm, including on tanned skin
- Pseudofolliculitis barbae
- Rejuvenation: tone, texture and diffuse redness
For dedicated laser hair removal: Lutronic Clarity II.
Discover all the technologies for treating vascular lesions distributed by Bioskin Italia.
DATASHEET
Technical specifications
| Laser sources | 532 nm KTP and 1064 nm Nd:YAG, solid state |
| Maximum pulse energy | 532 nm: 13.5 J 1064 nm: 100 J |
| Pulse duration | 532 nm: 0.3 to 40 ms 1064 nm: 0.1 to 60 ms |
| Pulse modes | Single pulse, sub-milli and sub-micro (Variable Sequential Pulsing) |
| Repetition rate | 0.5–10 Hz |
| Spot sizes | 2 to 16 mm, user-selectable; optical fibre delivery |
| Cooling | Cryogen, with the patented ICD (Intelligent Cooling Device) |
| Treatment control | IntelliTrak and real-time skin temperature sensing |
| Preset protocols | IntelliTone, Clear & Smooth, ART (Accelerated Rejuvenation with Tracking) |
| Dimensions | 41.8 × 95.0 × 100.8 cm (width × depth × height) |
| Weight | 120 kg |
| Power supply | 220–230 V AC, 50/60 Hz; 6.0 kVA power consumption; 30 A circuit breaker |
RESULTS, NOT WORDS
DERMAV: BEFORE AND AFTER


DermaV with Bioskin Italia
Bioskin Italia has been distributing aesthetic medicine devices since 1997. With DermaV we provide installation and clinical training at your practice or through training days at our partner reference medical centres, after-sales technical service and maintenance agreements, and marketing support to launch the treatment.
DermaV is also available through leasing or operational rental.
FAQ
Frequently asked questions about DermaV
What is the difference between 532 nm and 1064 nm?
532 nm works at the surface and is strongly absorbed by haemoglobin and melanin, so it suits fine vessels and pigmented spots. 1064 nm penetrates deeper and spares epidermal melanin, so it is indicated for larger vessels and darker skin types.
Which skin types can be treated?
1064 nm can be used on Fitzpatrick skin types I to VI, including tanned skin. 532 nm is used with more caution on darker skin, where melanin competes with the target.
Does the treatment require anaesthesia or downtime?
Usually not: cryogen cooling makes the treatment tolerable. Transient redness and mild swelling may appear after the session; patients can usually return to their normal activities straight away.
How many sessions are needed?
It depends on the lesion, its depth and the skin type. Facial telangiectasia often responds in a few sessions; port-wine stains and leg veins require more. Treatment protocols are covered during training.
Can I see DermaV in action before buying?
Yes: we can arrange a demonstration at your practice or at one of our partner reference medical centres, where training days are also held. Fill in the form below.